Effect of Span and MRD Configurations on Small Animal PET Image Quality and Quantitative Accuracy | ||
Journal of Biomedical Physics and Engineering | ||
مقالات آماده انتشار، اصلاح شده برای چاپ، انتشار آنلاین از تاریخ 20 خرداد 1404 اصل مقاله (2.45 M) | ||
نوع مقاله: Original Research | ||
نویسندگان | ||
Tahereh Zare1، 2؛ Mohammad Ghorbanzadeh2؛ Behnoosh Teimourian Fard2؛ Peyman Sheikhzadeh1، 3؛ Pardis Ghafarian4، 5؛ Sanaz Hariri Tabrizi2؛ Mohammad Hossein Farahani2؛ Mohammad Reza Ay* 1، 2 | ||
1Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran | ||
2Research Center for Molecular and Cellular Imaging (RCMCI), Advanced Medical Technologies and Equipment (AMTEI), Tehran University of Medical Sciences (TUMS), Tehran, Iran | ||
3Department of Nuclear Medicine, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, Tehran, Iran | ||
4Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD), Shahid Beheshti University of Medical Sciences, Tehran, Iran | ||
5PET/CT and Cyclotron Center, Masih Daneshvari Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran | ||
چکیده | ||
Background: Employing 2D rebinned sinograms in PET scanners has the potential to accelerate the overall reconstruction speed. Among the available rebinning techniques, Single-Slice Rebinning (SSRB) offers a computationally efficient approach. Objective: This study aimed to evaluate the influence of varying span and Maximum Ring Difference (MRD) parameters in SSRB on the image quality of the Xtrim PET scanner. Material and Methods: This Monte Carlo simulation study used a GATE-simulated Xtrim-PET scanner. 3D list-mode data were histogrammed into 576 sinograms, and SSRB was applied to generate 2D sinograms. Subsequently, Maximum-Likelihood Expectation-Maximization (MLEM) reconstruction was performed on the sinograms with different MRD and span. Image quality was assessed using image quality, rod, and uniform phantoms. Furthermore, axial resolution was evaluated using point sources. Results: Analysis of linear profiles in uniform phantom revealed a 2.6 mm inaccuracy in axial activity estimation when comparing spans of 21 and 7. Increased span and MRD lead to artifactual data in regions of high activity gradients, as observed in both uniform and rod phantoms. However, the Recovery Coefficient (RC) and Spilled-Over Ratio (SOR) remained unaffected. Concomitantly, increasing the span improved uniformity and reduced the coefficient of variation by 1.6% and 5.9%, respectively. Axial resolution remained largely unaffected by variations in span and MRD. Conclusion: The RC and SOR remain robust to variations in span and MRD. However, higher levels of axial data compression were associated with the introduction of axial artifacts. Additionally, axial resolution was unaffected by increases in span and MRD, likely due to the limited field of view of the Xtrim-PET scanner. | ||
کلیدواژهها | ||
Positron-Emission Tomography؛ Image Enhancement, Signal to-Noise Ratio؛ Rebinning؛ SSRB | ||